Benzene Separation Tubes 10-Pack for UltraRAE 3000+ by Honeywell RAE Systems

Area Monitoring
Gas Monitor Consumable Manufacturer: Honeywell RAE Systems
Benzene Separation Tubes 10-Pack for UltraRAE 3000+ by Honeywell RAE Systems
Benzene Separation Tubes 10-Pack for UltraRAE 3000+ by Honeywell RAE Systems
Benzene Separation Tubes 10-Pack for UltraRAE 3000+ by Honeywell RAE Systems

RAE-Sep Benzene Separation Tubes for the Honeywell RAE Systems UltraRAE 3000+ monitor. These disposable tubes chemically separate benzene from other interfering VOC compounds, enabling accurate and selective benzene measurement in complex environments.

  • Compatibility: UltraRAE 3000+ only
  • Detection Range: 0.1 to 1,000 ppm
  • Quantity: 10 tubes per box

Call for availability and lead times prior to ordering.

Choose Your Option

Rent for flexible short-term use or purchase to own

Rent
Purchase
Need Help Choosing?
Our rental specialists can help you select the right equipment for your specific application.
Call 866-736-8347
1
Choose Purchase Option
Rent short-term or buy to own
Required
No dates set, use the "Set Dates" button above to select your rental period.
Important: These dates apply to your entire cart. Changing dates will affect pricing for all rental items in your cart.
Used Equipment: This equipment has been previously used and tested. It may show signs of wear but is fully functional and inspected.

Your Configuration

Benzene Separation Tubes 10-Pack for UltraRAE 3000+ by Honeywell RAE Systems
× $255.00
$255.00 Select dates
Subtotal:
$255.00
Select dates
Technician verified and tested instrumentation before every rental
Expert support available by phone and email
Fast delivery and flexible rental terms

About This Product

RAE-Sep Benzene Separation Tubes for the Honeywell RAE Systems UltraRAE 3000+ monitor. These disposable tubes chemically separate benzene from other interfering VOC compounds, enabling accurate and selective benzene measurement in complex environments.

  • Compatibility: UltraRAE 3000+ only
  • Detection Range: 0.1 to 1,000 ppm
  • Quantity: 10 tubes per box

Call for availability and lead times prior to ordering.

Exposure Hazards This Equipment Detects

Volatile Organic Compounds (VOCs)

CAS: N/A
Learn More
PEL TWA
N/A
REL TWA
N/A
TLV TWA
N/A

Benzene (C₆H₆)

CAS: 71-43-2
Learn More
PEL TWA
1 ppm
REL TWA
Ca 0.1 ppm
TLV TWA
0.02 ppm

Features & Capabilities

Monitoring Scope: Area Monitoring

Equipment Specifications

  • Target Compound:Benzene (C₆H₆)
  • Detection Range:0.1 to 1000 ppmv
  • Sample Time:60 seconds at 15–30°C (temperature dependent)
  • Sample Volume:360 mL (standard at 15–30°C)
  • Color Change:Yellow → Brown → Green (visual indication of breakthrough)
  • Operating Temperature Range:2°C to 40°C (36°F to 104°F)
  • Humidity Tolerance:0% to 95% RH (no effect on readings)
  • Storage Life:1 year unopened, 5–25°C (preferably refrigerated)
  • Cross-Sensitivity:Minimal interference from common VOCs; e.g., toluene = 0.1, xylene = 0.0, styrene = 0.0
  • Tube Contents:~3 mg of chromium compounds (dispose per regulations)

Manufacturer Resources

Documents

User manuals, safety guides, and technical documentation from the manufacturer:

Videos

No manufacturer videos are currently available for this product. Check back later or contact us for more information.

Ask a Question

Have a question about this product? Fill out the form below and our team will get back to you shortly.

Need Help Choosing the Right Equipment?

Our customer success specialists are here to help you select the perfect equipment for your project