LT300 by Extech - Light Meter
LT300 by Extech - Light Meter
Note: This unit is calibrated every 2 years. NIST calibration is valid for 1 year. Contact RAECO Rents to confirm availability if required.
The Extech LT300 is a digital light meter with a measurement range of 1 to 200,000 Lux (0.1 to 20,000 Foot-candles), using a detachable CIE photopic-response photo sensor on a 12 in. coiled cable (expandable to 24 in.). Built-in functions include Max/Min recording, Peak Hold (captures pulses as fast as 10 µs), Data Hold, and Relative mode for differential comparisons. Large backlit LCD with bargraph display. Suitable for OSHA lighting compliance checks, illumination audits, and energy efficiency assessments in offices, factories, hospitals, cleanrooms, and schools.
Choose Your Option
Rent for flexible short-term use or purchase to own
Exposure Hazards This Equipment Detects
Illuminance (lux / fc)
Features & Capabilities
| Calibration Certificate Available: | ✓ Yes |
| Monitoring Scope: | Area Monitoring |
Equipment Specifications
| Measurement Type: | Illuminance (lux / foot-candles) |
| Detection Method / Sensor Type: | Silicon photodiode with spectral response filter, on a coiled remote sensor cable |
| Measurement Range: | 0 to 200,000 lux in 5 ranges (40, 400, 4,000, 40,000 and 200,000 lux); 0 to 20,000 fc in 4 ranges (40, 400, 4,000 and 20,000 fc) |
| Spectral Range: | CIE photopic (human eye) response; spectral accuracy f1′ ≤ 6% |
| Resolution: | 0.01 lux / 0.01 fc on the lowest range |
| Accuracy: | ±(5% of reading + 0.5% of full scale); repeatability ±2%; temperature coefficient ±0.1% per °C |
| Sampling Rate: | 1.3 readings per second (digital display); 13.3 per second (bargraph) |
| Additional Features: | MAX/MIN, Data Hold, Relative and Peak modes; backlit display; tilt stand and tripod mount; protective sensor cap |
| Peak Capture: | Light pulses down to 10 µs (PMAX / PMIN) |
| Calibration Standard: | Calibration certificate included. If your program requires a NIST calibration within the last year, call us before ordering |
| Recalibration Interval: | Every 2 years (Extech does not set a calibration interval) |
| Connectivity: | No data logging or PC connection |
| Display: | 3-3/4 digit (3999 count) LCD with analog bargraph and backlight; OL over-range indication |
| Operating Temperature Range: | 32°F to 104°F (0°C to 40°C), below 80% RH |
| Storage Temperature Range: | 14°F to 140°F (-10°C to 50°C), below 80% RH |
| Battery Run Time: | 9V battery; run time not specified by Extech |
| Size / Weight: | 6.5 x 3.0 x 1.7 in (165 x 76 x 43 mm); 14.2 oz (403 g) with battery |
Kit Contents
Manufacturer Resources
Documents
User manuals, safety guides, and technical documentation from the manufacturer:
Videos
No manufacturer videos are currently available for this product. Check back later or contact us for more information.
Ask a Question
Have a question about this product? Fill out the form below and our team will get back to you shortly.
- Translation missing: en.general.accessibility.refresh_page
- Translation missing: en.general.accessibility.new_window